Setting Up the XP3000 for Your Application
4-4
2
Select syringe size.
Determine your volume and flowrate requirements. Select a syringe that
accommodates the smallest and largest volumes to be dispensed without refill, as
well as the desired flowrate (see Table 4-1). While smaller syringes allow better
accuracy and precision, a larger syringe allows more aliquots when multiple
aspirations or multiple dispenses are required, and they allow better breakoff and
longer seal life.
Table 4-1. Flowrate Ranges
Syringe Size
Minimum Flowrate
(mL/min)
Maximum Flowrate
(mL/min)
50
µ
L
0.0025
2.6
100
µ
L
0.0050
5.23
250
µ
L
0.012
13.1
500
µ
L
0.025
26.2
1 mL
0.050
52.3
5 mL
0.25
261.6
3
Select tubing.
In tubing selection, the general rule is that smaller syringes work best with smaller
I.D. tubing and larger syringes with larger I.D. tubing. Most XP 3000 valve styles
have an internal I.D. of 0.059" (approx. 1/16"). For aspirate/dispense tubing a
thermal-drawn tip or tapered tip is most common, providing good breakoff and
excellent accuracy and precision for most applications. A necked-down tip may be
used when aspirating very small volumes of sample, i.e., 1 - 5
µ
L. A blunt-cut tip is
better suited for large volume applications. For tubing recommendations, see Table
4-2; for a description of the various types of tubing, see Appendix A, “Ordering
Information.”
Table 4-2. Tubing Recommendations
Syringe Size
Aspirate/Dispense Tubing P/N
Reagent Tubing P/N
50
µ
L, 100
µ
L, 250
µ
L
5133
721370
5723
*
500
µ
L, 1 mL, 2.5 mL
5133
4609
720595
5729
720597
721370
5 mL
4333
720592
720595
721370
Summary of Contents for XP 3000
Page 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Page 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Page 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Page 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Page 134: ......