Software Communication
3-46
Table 3-8. Error Codes and ASCII and Hexadecimal Values
Status Byte
Hex # if
Bit 5 =
Dec # if
Bit 5 =
Error
Code
7 6 5 4 3 2 1 0
0
or 1
0
or 1
Number
Error
0 1 X 0 0 0 0 0
40h
60h
64
96
0
No Error
0 1 X 0 0 0 0 1
41h
61h
65
97
1
Initialization
0 1 X 0 0 0 1 0
42h
62h
66
98
2
Invalid Command
0 1 X 0 0 0 1 1
43h
63h
67
99
3
Invalid Operand
0 1 X 0 0 1 0 0
44h
64h
68
100
4
Invalid Command Seq.
0 1 X 0 0 1 0 1
45h
65h
69
101
5
Fluid Detection
0 1 X 0 0 1 1 0
46h
66h
70
102
6
EEPROM Failure
0 1 X 0 0 1 1 1
47h
67h
71
103
7
Device not Initialized
0 1 X 0 1 0 0 1
49h
69h
73
105
9
Plunger Overload
0 1 X 0 1 0 1 0
4Ah
6Ah
74
106
10
Valve Overload
0 1 X 0 1 0 1 1
4Bh
6Bh
75
107
11
Plunger Move Not
Allowed
0 1 X 0 1 1 1 1
4Fh
6Fh
79
111
15
Command Overflow
Error Reporting Examples
[A4000R]
Does not return an error immediately after the
command, but when queried ([Q] command),
returns an “Invalid Parameter” error.
[A3000A3500R]
Moves to position 3000, then stops. A [Q]
command returns an error.
[E2000R]
Returns an invalid command error immediately.
The pump status is “Not Busy.”
[A3000E2000R]
Returns an invalid command error immediately.
The pump is “Not Busy.”
Valve in Bypass [A1000R]
Does not return an error immediately, but when
queried ([Q] command), returns a “Plunger Move
Not Allowed” error.
Summary of Contents for XP 3000
Page 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Page 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Page 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Page 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Page 134: ......