Software Communication
3-36
L <n> S
ET
S
LOPE
During the beginning and end of a move, the plunger ramps up and down to top speed.
The ramp is programmed using the Slope command. It is calculated as <n> x 2.5
kHz/sec. The syntax for this command is:
[L<n>], where <n> = 1..20 (14 is the default)
The corresponding slopes in kHz/sec are listed below.
Slope Code
kHz/Sec
1
2.5
2
5.0
3
7.5
4
10.0
5
12.5
6
15.0
7
17.5
8
20.0
9
22.5
10
25.0
11
27.5
12
30.0
13
32.5
14
35.0
15
37.5
16
40.0
17
42.5
18
45.0
19
47.5
20
50.0
v
<n> S
TART
V
ELOCITY
The [v] command sets the velocity at which the plunger begins its movement. The
plunger will then ramp up (slope) to the top velocity. The start velocity should always be
less than the top velocity. The syntax for this command is:
[v<n>], where <n> = 50..1000 Hz
(900 is the default; 901 is the default on the 3-port distribution valve)