43
5-3-5. Option Function
F66
Function
Select Option Type
VFD Display
DESCRIPTION
F66 0
Do not use
F66 1
BCD Out (Option-3)
F66 2
Analog Out (Option-4, 5)
F66 3
Current Loop (Option-1)
Set Value
(0 ~ 5)
F66 4
BCD In (Option-6)
F68
Function
Output Current of Voltage at Display Zero
VFD Display
DESCRIPTION
L 00000
0 mA, 0V
L 04000
4.000 mA, 2V
Set Value
(0 ~ 24000)
L 04015
4.015 mA, 2.007V
F69
Function
Output Current of Voltage at Full Scale
VFD Display
DESCRIPTION
H 10000
10 mA, 4.16V
H 20000
20.000 mA, 8.33V
Set Value
(0 ~ 24000)
H 24000
24.000 mA, 10V
F72
Function
Select Output Logic - BCD Out
VFD Display
DESCRIPTION
F72 0
Positive Logic
Set Value
(0, 1)
F72 1
Negative Logic
F73
Function
Set Point Inpu
VFD Display
DESCRIPTION
F73 0
Disable external Set point input
Set Value
(0, 1)
F73 1
Enable external Set point input