17
(4) Front slide S/W usage
If you open the front cover of the indicator, you will see the Dip switch.
You can enter the related Mode to the state of the SW.
□
DIP Switch : Mode Table
□
Test Mode
▣
If you turn off all of the DIP swtiches(0,0), you can move into the Test Mode then.
▣
If you turn on all of the DIP swtiches(1.1) after the Test, you can return to the Weighing Mode
then.
□
Calibration Mode
▣
If you turn on only the DIP switch 1 (0, 1), you can move into the Calibration Mode then.
▣
If you turn on all of the DIP switches (1,1) after the Calibration, you can return to the
Weighing Mode then..
□
Set Mode
▣
If you turn on only the DIP switch 2(1, 0), you can move into the Set Mode then.
▣
If you turn on all of the DIP switches(1,1) after the Calibration, you can return to the
Weighing Mode then.
DIP S/W
Mode
0 0
Test Mode (TEST)
0 1
Calibration Mode (CAL)
1 0
Set Mode(SET)
1 1
Weighing Mode (WEIGHING)