2. Installation
71
Precautions for connecting the server (reference)
Equipment Restrictions
on
connections
Kodak
Miil
Transmitting the 0019 shadow group causes a failure, and the group is not received
properly with the default. Its reception is enabled by setting the strictValidation
parameter to Off in Miil.
Fujitsu
Dr. ABLE
y
Transmission is currently performed with the “1 study multi series/1 series 1
image” setting. However, since a multi format is used for the screen displays for
each series under the Dr.ABLE specifications, the switching operations are a
hassle. The user will find it more convenient if it is at all possible to change the
setting to “1 study 1 series/1 series multi image.” (These unusual data
specifications were requested with the full understanding of their unusualness.)
y
With DICOM, the body parts (such as the abdomen and head) and their directions
(such as PA and AP) belong to the series information. It therefore follows that a
different series is required for a different body part or body part direction. This
aspect is restricted by the DICOM standard rather than by the installation and
other steps taken by us. To put it the other way around, multiple images with
different body parts and their directions cannot be put together as a series. To
remedy this problem, devising a way of enabling the viewers to reference
different series at the same time at some future point in time will be helpful.
Hitachi
y
Transmission is currently performed with the “1 study multi series/1 series 1
image” setting. However, since a multi format is used for the screen displays for
each series, the switching operations are a hassle. The user will find it more
convenient if it is at all possible to change the setting to “1 study 1 series/1 series
multi image.”
y
This problem arises with the Fujitsu equipment as well. Refer to the section on
Fujitsu.
send_image error return values and log output (reference)
Example of
error
Return value and log output
Successful Return
value:
0x00
No log output at this time
Invalid parameter
exists.
Return value: 0x00000001
CXDI interpretation:
DICOM connect error
(CCRTRANS_ERR_DICOMPARAM)
See below for the log output at this time:
-------------------------------------------------------------------------------------
-t
Set called AE title to title in Association RQ
-v
Place DUL and SRV facilities in verbose mode
node
Node name for network connection
port
TCP / IP port number of server application
image
A list of one or more images to send
-------------------------------------------------------------------------------------
Server fails to start. Return value: 0x00180012
CXDI interpretation:
DICOM connection error
(CCRTRANS_ERR_NOT_CONNECT)
See below for the log output at this time:
-------------------------------------------------------------------------------------
18-132600[127]ERR:
60012 TCP Initialization Error: Invalid argument
18-132600[127]ERR:
130012 Peer aborted Association (or never
connected)
18-132600[127]ERR:
180012 Failed to establish association
-------------------------------------------------------------------------------------
As a result of
providing the
server with a
command request,
an error was
returned as
response.
Return value: 0x10
CXDI interpretation: DICOM response error (CCRTRANS_ERR_RESP)
See below for the log output at this time:
-------------------------------------------------------------------------------------
(To be determined)
-------------------------------------------------------------------------------------
Summary of Contents for CXDI-60G
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...