2. Installation
30
7.3.3 Rating and Characteristics for Relay and Photo Coupler (on PWB-
60XRAY)
(1) RL1 (Power Relay/Plug-in terminal type)
1)
Rating (Operational coil)
Rated
curre
nt
Coil
resistance
Coil Inductance
(mH)
Pick-up
voltage
Dropout
voltage
Maximu
m voltage
Power
consumption
Rated
voltage
(V)
(mA) (
Ω
)
Armature
OFF
Armature
ON
(V) (V) (V) (mW)
DC5V 72 69.4
69.5
86.0
below
80%
below
10%
135%
(at 50)
Approx 360
2) Rating (Switch/Contact)
types Single
stable
arrangement 2
Form
C
Contact material
Au-clad AgNi type
Relating capacity
AC250 5A
DC30V 5A
Max. switching power
1250VA 150W
Max. switching voltage
250V AC
Max. switching current
5A
Min. switching capacity
100
µA 1V DC
3) Characteristics
Item Content
Operate time
Max. 20ms
Reset time
Max. 10ms
Mechanical 18,000
times/hour
Maximum open/close
frequency
Rated load
1,800 times/hour
Between coil
contacts
2,000 Vrms
Withstand voltage
Between same poles
1,000 Vrms
Mechanical 5x10
7
times
Life
Electrical
10
5
at 5A 250V AC
5x10
5
at 5A 30V DC
Ambient temperature
-40
°
C
to +70
°
C
(no freezing nor condensation)
Maximum operating frequency
50 times/Sec.
Summary of Contents for CXDI-60G
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...