Examples of parameters used with different makers and types of printers (reference)
Name Default
Parameters
Significance
Kodak MLP190
-A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
80 (entered in pixelPitch
member)
14 x 17
4096 (entered in W member)
5120 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing: normal
•
Maximum density: 3.20
•
Curve shape 0 (density linear) as Config
Info
Kodak KE
Kodak Print Spooler
Model 100
-A 320 -T NO -M CUBIC –m
NORMAL -S CS000
(entered in param member)
79 (entered in pixelPitch
member)
14 x 17
4090 (entered in W member)
5120 (entered in H member)
11 x 14
3194 (entered in W member)
4096 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing: normal
•
Maximum density: 3.20
•
Curve shape 0 (density linear) as Config
Info
Agfa DryStar 3000
-A 320 -T NO -M CUBIC -m
140 -S
“PERCEPTION_LUT=200”
(entered in param member)
80 (entered in pixelPitch
member)
14x17
4256 (entered in W member)
5174 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing: slightly sharp (edges
emphasized)
•
Maximum density: 3.20
•
S
“PERCEPTION_LUT=200(LINEAR)”
(If the output fails to be linear with
“LINEAR”, on-site adjustments with the
printer manufacturer must be performed.
Kodak Imation DryView
8700 + Pacs LINK IMN
9410
-A 310 -T NO -M CUBIC -m
? -S LUT=0?, 2? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing must be adjusted at the user’s
site.
•
Maximum density: 3.10
•
S LUT = m, n is designated as the
Config Info but m and n are adjusted on-
site by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
Kodak Imation DryView
8700 + GW
-A 310 -T NO -M CUBIC -m
? -S LUT=?, ? (?: Site
Dependent)
(entered in param member)
78 (entered in pixelPitch
member)
14x17
4096 (entered in W member)
5220 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Smoothing must be adjusted at the user’s
site.
•
Maximum density: 3.10
•
S LUT = m, n is designated as the
Config Info but m and n are adjusted on-
site by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
Kodak Imation DryView
8700+8800
-A 320 -T NO -M CUBIC –S
“LUT=m, n”-m (on-site
adjustment)
(entered in param member)
78 (entered in pixelPitch
member)
4096 (entered in W member)
5220 (entered in H member)
•
Trimming OFF
•
Cubic spline interpolation
•
Maximum density: 3.20
•
S LUT = m, n is designated as the
Config Info but m and n are adjusted on-
site by the Kodak service engineer.
Basically, adjustment is performed to
achieve a linear output.
Summary of Contents for CXDI-60G
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...