4. Repair Guide
24
3) Tilt the grid unit toward the imaging unit side, and fit the two anchoring hooks (see
detail C) of the lock mechanism into the two cavities (see detail D) on the handle sides
of the imaging unit.
* Check that the anchoring hooks automatically fit into the cavities and lock.
4.3 Checking the release of the installed grid unit
1) Place the imaging unit with the installed grid unit on a flat surface, and slide the two
slide switches on the outer side of the lock mechanism toward the outside. This will
cause the hooks fitted into the cavities of the imaging unit to be pulled up in tandem,
and the lock to be released.
2) With the slide switches still slid, hold both sides of the grid frame, and separate it from
the handle side of the imaging unit.
3) Separate the two tabs on the grid frame from the two cavities of the imaging unit, and
check that they disengage.
Grid frame
Imaging unit
Summary of Contents for CXDI-60G
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...