1. General
3
3 System Diagram
3.1 Standalone System
Standalone System Block Diagram
AC or DC IN
Protective
Grounding
1st / 2nd
AC250V
Insulation with
reinforcement
Power Supply
AC250V
Basic Insulation
Imaging Unit
Grid Uni
t
Remote Switch
X-ray Generator
(601Compatible)
FC-24VE/FC-E21A
or
Recommended general computer
Patient Circumstances or Medical Room
Power Box
SELV
*
Control/Signal
Control
Control/Signal
Control/Signal
Ethernet
* SELV = Safety Extra-Low Voltage
CXDI-60G/60C
General HUB
(1)
(2)
Connector
3.2 Total System
It can be connect to (1) and (2) of the system where the existing products have already been
connected. Extend the ethernet port by general switching HUB or ethernet card. The maximum
number of connections is limited to four by the control software specification.
Total System Block Diagram (Example)
AC or DC IN
Protective
Grounding
1st/2nd
AC250V
Insulation
with reinforcement
Power Supply
AC250V
Basic Insulation
Imaging Unit
G
rid
Un
it
Remote Switch
X-ray generator
(601 Compatible)
FC-24VE/FC-E21A
or
Recommended general computer
Patient Circumstances or Medical Room
Power Box
SELV
*
Control/Signal
Control
Control/Signal
Control/Signal
Ethernet
*SELV = Safety Extra-Low Voltage
j
CXDI-60G/60C
General HUB
Power Box
CXDI-40E
Imaging Unit
Control/Signal
Ethernet
Power Box
CXDI-50
Imaging Unit
Control/Signal
Ethernet
Ethernet
Image Examination WS
Image File Equipment
Printer etc.
Connector
Summary of Contents for CXDI-60G
Page 23: ...2 Installation 7 2 CXDI System II assembly package Accessories box Power box ...
Page 115: ...2 Installation Manual 99 2 Power Box Mass 3 7 Kg Unit mm Dimension tolerance 1 Scale 1 N ...
Page 161: ...4a Canon Inc PWB 60 X RAY 1 2 SCHEMATIC DIAGRAM Rev 01 ...
Page 162: ...4b Canon Inc PWB 60 X RAY 2 2 SCHEMATIC DIAGRAM Rev 01 ...