AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
6.1.1 Pipeline Mode
When the CRBasic program first starts running, the information specified in
the AVW200( ) instruction is sent to the attached AVW200 interface module
via the communication port and PakBus address specified in the instruction.
Along with the instruction's parameter information, the datalogger also sends
its clock information. For every execution interval, the datalogger clock and
measured data are exchanged with the attached AVW200 interface module.
The AVW200 interface module then determines how much time previous to
the datalogger’s interval it should start making the vibrating wire measurement,
so that it completes the measurement series before the datalogger's execution
interval.
A returned “Result” value (the first parameter in the AVW200( ) instruction) of
zero indicates successful communication and tells us that the data values have
been stored in the destination variable. If there is a failure in communication,
the “Result” value increments for each failure and the destination values are
filled with NAN's. The datalogger will retry communications three times
before returning a failed communication or incrementing the result (retries are
every 3 seconds or greater depending on the radio-power-cycle configuration).
A negative value returned for the “Result” variable indicates status information
(e.g., a -3 indicates the AVW200 has not made the first measurement; -4
indicates that no parameter information is available). Multiple AVW200()
instructions can use the same “Result” variable. There is no reason to have
different “Result” variables for a given communication port in the pipeline
mode.
In the Pipeline mode of operation the AVW200( ) instruction
cannot be in a conditional statement or in a subroutine.
NOTE
Read more! An example program of the pipeline mode is provided in
Section 7.3.
6.1.2 Sequential Mode
In sequential mode, during each execution interval, the CRBasic program
executes each instruction in sequence (i.e., the first AVW200( ) instruction is
executed through to completion, followed by the second instruction…). The
first AVW200( ) instruction will communicate with the attached AVW200
interface module and tell the AVW200 to make the measurements. The
datalogger will then wait at the AVW200( ) instruction until all of the
measurements are returned for that first instruction, or until the instruction
times out (time out is two times the number of reps). If an instruction times
out, it will make three additional attempts before advancing to the next
instruction. The next AVW200 instruction is then executed.
The result codes returned via sequential mode function the same as they do in
the pipeline mode. However, in sequential mode, it is a good idea to have
different result code variables for each AVW200( ) instruction. This is because
if communication was disconnected between two AVW200( ) instructions, then
data collected via the first instruction would correctly be stored into destination
variables for that instruction. But, because of the break in communication, the
destination variables for the second instruction would be filled with NANs and
the result code would increment (indicating a failed communication). With
41
Summary of Contents for AVW200 series
Page 2: ......
Page 78: ...Appendix C Antennas Antenna Cables and Surge Protectors for the AVW206 AVW211 and AVW216 C 8...
Page 86: ...Appendix E Status Table E 4...
Page 104: ...Appendix H Additional Programming Examples H 12...
Page 110: ...Appendix I Using MD485 Multidrop Modems with AVW200 Interfaces I 6...
Page 111: ......