AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
5.4 Send OS
For most applications, Campbell Scientific does not anticipate that it will be
necessary to download a new operating system to the AVW200. However, if a
new operating system (OS) is required, in order to send a new OS to the
AVW200 you will need Device Configurator (DevConfig) 1.10 or greater.
First connect the RS-232 port of the AVW200 to a serial port on your computer
using a 9-pin serial cable and follow the steps below.
1.
Remove power from the AVW200.
2.
Open DevConfig.
3.
Highlight the AVW200 in the list of devices which appears in the left-hand
portion of the window.
4.
Select the COM port to which the AVW200 is connected from the drop
down list box at the bottom left of the window.
5.
Click the Send OS tab and follow the directions on the screen (Fig. 5.4-1).
FIGURE 5.4-1. Sending an Operating System using DevConfig