92
of 216
6.3.1.6 Strain gage sensors
You can connect strain gage sensors to connector B.
Diagram 38: Channel settings for strain gage sensors - page 1
Note:
A 5 VDC excitation voltage is used for strain gages.
Parameters in the "Y-axis Socket B Strain gage" menu (M22) – page 1
Sensitivity
0.02 to 100 mV/V
Enter the strain gage sensitivity
This value is used solely for calculating and displaying
the electrical output level.
This parameter has no relevance to the internal channel
settings.
Input range
2, 4, 10, 20, 40 mV/V
Input range selection for the strain gage
In order to be able to use 100 % of the measurement
range of the connected sensor, the selected input range
must be ≥ the sensor sensitivity.
Level (el.)
Value in %
Shows the electrical output level from the measurement
channel
Filter
Off, 5, 10, 25, 50, 100,
200, 400, 800 Hz
Filter setting for the measurement channel
(Default value = 50 Hz)
Units
aaaa, bbbb, cccc, mm,
N, kN, Nm, Ncm, deg,
bar, V, s, ms
Units for the measurement channel
You can select the units from a list.
Alternatively, you can also specify user-defined units.
Note:
For further information on
[User-defined units]
[Tare Setup]
and
[Read TEDS
data
], please
refer to the relevant subsections below.