89
of 216
7
Enable the checkbox "Tare at meas. start“
and use the keypad to enter the default tare
value ("Standard value for tare").
If you enable the "OUT_WARNING_TARE"
checkbox, then the "OUT_WARNING_
TARE" control output will be set if the Tare
warning limit is exceeded.
8
Tap to return to the menu "X/Y-axis Socket A Stand. signal".
Parameters in the "X/Y-axis Stand. signal Taring“ menu (M62)
Tare at meas. start
On / Off
Enable/disable
Enabled:
At every start of a measurement, the channel is
automatically tared to the default tare value
("Standard value for tare").
Standard value for
tare
<value input>
A typical tare value is 0.0.
OUT_WARNING_TARE
On / Off
Enable/disable
Enabled:
If the signal to be tared exceeds the "Tare warning
limit", the "OUT_WARNING_TARE" control output
is set.
Note:
This may also indicate that the sensor is
faulty.
Tare warning limit
--- / 1 to 20 %
The "Tare warning limit" can be set between 1 %
and 20 % with respect to the actual measurement
range of the input channel (regardless of the
current scaling).
Note:
Please note when using sensors with a
high zero point offset, the function “Tare
warning limit” can only be used in a
limited extent or not at all.