VÅNTEC-1 Detector User Manual
Hardware Installation for the D8 ADVANCE
M88-E01072
4 - 13
4.7.2 Connection and Cable Wiring
The cabling differs in one way compared to that
in Figure 4.6:
1.
The motor clock cable (A17-D42) has to be
connected between the 9-pin socket STEP/
DIR of the Detector Controller or external
FDC and either:
•
For D8 systems, the X5 9-pin socket on the
B231 2-axis indexer board, or
•
For D4 systems, the socket on the bracket
(X11) of the B232 4-axis indexer board.
NOTE: The clock signal cable can be one of two
cables, one straight and one Y-shaped. If you
are running the system with the AIB2G, NEVER
use the Y-shaped adapter cable (A17-D43) in
any case!
After connecting the cables, switch on the
detector controller and D8/D4 system.
Figure 4.8 - Clock signal cable (A17-D42)
Figure 4.9 - Y-shaped adapter cable (A17-D43).
NOTE: Use only for first-generation indexer
boards (AIB) B104/B105.
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...