M88-E01072
8 - 1
8. Alignment
Peak position is determined during calibration
(see Section 7), not during alignment. The align-
ment optimizes the resolution of the detector.
The resolution is determined by finding the best
possible FWHM (Full Width Half Maximum). The
resolution is dependent on the measuring circle
and the alignment of the detector mount.
8.1
Alignment of the VÅNTEC-1
Detector and Mount
Complete a coarse alignment of the 2 DOF
mount by running repeated PSD fixed scans of
NIST SRM 1976 or Quartz. For a quick assess-
ment of the alignment, compare the intensities
of the main (104) peak to the valley between
K
α
1
and K
α
2
(see the example in Figure 8.1).
[Example: The first scan’s intensity of the valley
is 4,000 cps. The main peak’s intensity is 8,000
cps. This represents a ratio of 2 (4,000:8,000 =
1:2)]. A ratio of 3.0 to 4.0 is desired. Once the
max ratio is achieved, proceed to Section 8.2.
1.
Run a PSD fixed scan. Use a standard ref-
erence material of the NIST SRM 1976
Corundum plate (104) or Quartz (101)
reflection. Use the parameters in Table 8.1
for SRM 1976.
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...