VÅNTEC-1 Detector User Manual
Hardware Installation for the D8 ADVANCE
M88-E01072
4 - 31
4.8.4 Mount and Align the Radial Soller Slit
Use the following tools for this procedure:
•
NIST SRM 1976 Corundum plate (for reflec-
tion) or NIST SRM 640c Silicon powder (for
transmission)
•
Torx key set
•
Allen key set
Mount the radial Soller slit:
1.
Take off the k
β
filter.
2.
Attach the radial Soller slit onto the VÅN-
TEC-1 optics block using the two anti-scat-
ter slit screw holes shown in Figure 3.9.
3.
Remove the anti-scatter slit assembly of the
detector optics and mount the radial Soller
slits with the two M3 screws. The alignment
pin of the detector optics block has to be
visible.
Figure 4.31 - Installation of the radial Soller slit
Fixing
screws
Alignment pin
Clamping
screws
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...