Hardware Installation for the D4 ENDEAVOR
VÅNTEC-1 Detector User Manual
5 - 8
M88-E01072
9.
Mount the actuator for the beam stop colli-
sion switch using a T-nut in the groove ring
(see Figure 5.10). The position of the actua-
tor is defined by the extreme low position of
the alignment screw and shows the highest
position of the primary beam stop (about 6°
2-theta).
Figure 5.9 shows the actuator in the
extreme high position of the alignment
screw.
Figure 5.10 - Mounting of the actuator bracket for beam stop
collision switch
10. Mount the collision switch to the detector
mount. Place the color-coded banana plugs
into the sockets on the goniometer side
panel labeled “Collision switch.” Do not
switch the colored plugs and sockets (see
Figure 5.11).
Figure 5.11 - Cabling of the collision switch
11. Connect the plug X672 to the cable for the
switch monitoring in D4TOOLS.
Actuator/
bracket
Mount
positions
Fixing of
actuator
Plugs of
collision
switch
Connector for the
switch monitor
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...