System Description
VÅNTEC-1 Detector User Manual
3 - 18
M88-E01072
3.5
Counterbalance
The counterbalance is located on the back of
the goniometer. Figure 3.16 shows an example
of a
counterbalance on a
θ/θ
goniometer. The
addition of a counterbalance provides smooth
and flawless movement of the goniometer. The
position of the counterbalance is based on the
weight of the detector and optics, the measuring
circle diameter, and the height of the sample
stage. In order to ensure proper weight distribu-
tion, please use the tables below to select the
correct position of the counterbalance for the
detector system. Table 3.3 refers to the posi-
tions shown in Figure 3.17.
Figure 3.16 - Counterbalance on the goniometer
Stage Height (mm)
150 214 258
Circle Diameter (mm)
435
500
600
Handle
-4
-1
1
Weight
B
B
B
Table 3.3 – Position of the handle and weight for the
counterbalance
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...