VÅNTEC-1 Detector User Manual
Alignment
M88-E01072
8 - 3
3.
Adjust the yaw “Y” of the 2 DOF. The yaw
adjustment uses opposing screws. To make
an adjustment, first loosen the opposing
screw and then tighten the adjusting screw
(see Figure 8.2).
Figure 8.2 - Yaw “Y” and roll “Z” adjustments
4.
Re-run the scan and compare the ratio to
the baseline scan. If the ratio increases,
continue to adjust in the direction first
attempted. If the ratio decreases, reverse
the direction of movement and repeat step 3
until the maximum ratio is achieved. Use the
ratio to check the alignment. Intensity of the
valley alone may vary.
5.
Repeat step 3 and step 4 for the roll “Z” of
the 2 DOF until a ratio of 3.5 to 4.0 is
achieved. Noticeable separation can be
seen between the K
α
1
and K
α
2
peaks when
alignment is correct. Save the final scan for
future reference.
roll “Z”
yaw “Y”
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...