M88-E01072
7 - 1
7. Calibration
Prior to calibration, the system must be aligned
using the scintillation detector.
Each detector and the associated electronics
are unique. A calibration process is used in con-
junction with standard reference materials to
determine a zero offset for the resolution in
degrees. This value represents the angular dif-
ference between the zero angle of 2-theta and
the zero channel of the PSD. The calibration
program will refine the values and record them
in the Configuration program.
7.1
Positional Calibration
1.
Manually open all window slits as wide as
possible.
2.
Ensure that the proper K
β
filter is in place.
3.
Make sure that no Debye slit is in place.
4.
Ensure that the reference material is
mounted (quartz or NIST 1976) in the sam-
ple stage.
5.
Set the goniometer to the parameters in
Table 7.1 or Table 7.2 using DIFFRAC
plus
COMMANDER.
NOTE: Do not close DIFFRAC
plus
COM-
MANDER. Instead, minimize the window.
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...