7 - 3
7.2
PSD Calibration
7.2.1 Create Files
1.
Open XRD Wizard from C:\Diffplus\XRD-
Wizard.exe.
2.
Select File and New to display the file types
available. This window creates the DQL
used for calibration.
3.
Select PSDCalib from the new menu files.
Click OK (see Figure 7.1).
Figure 7.1 - Wizard > File > New
4.
To create the DQL file, input the calibration
parameters and click OK. The number of
positions must be nine or greater. When-
ever possible, select rotation ON and set
the speed at 15 rpm.
5.
Determine the theoretical 2-theta values
using Table 7.1 and Table 7.2. Input the val-
ues into the Calibration Parameters screen
(see Figure 7.2).
6.
Input a step size of 1.
7.
Save the DQL file as Calib.DQL and exit
XRD Wizard.
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...