VÅNTEC-1 Detector User Manual
Hardware Installation for the D4 ENDEAVOR
M88-E01072
5 - 7
7.
Drive the sample holder to 30° and mount
the beam stop for the D4 (see Figure 5.8).
Place a T-nut into the groove ring of the
goniometer and tighten the fixing screw.
Ensure that the two additional pins also fit
into the groove ring.
NOTE: Do not lose the T-nut in the groove—put
a tissue into the groove to prevent this.
Figure 5.8 - Mount the primary beam stop
8.
Screw the alignment screw completely out
(see Figure 5.9). Put the alignment pin into
the goniometer ring and place the beam
stop so the alignment screw hits the align-
ment pin.
Figure 5.9 - Details of the primary beam stop—showing the
extreme low position of the beam stop
Primary
beam
stop
Alignment
screw
Linear
guidance
Alignment
pin
Fixing
screw
Positioning
pins
Summary of Contents for VANTEC-1
Page 1: ...V NTEC 1 Detector USER MANUAL M88 E01072 5 05 BRUKER ADVANCED X RAY SOLUTIONS...
Page 2: ......
Page 3: ...V NTEC 1 Detector User Manual BRUKER ADVANCED X RAY SOLUTIONS M88 E01072 5 05...
Page 34: ...System Description V NTEC 1 Detector User Manual 3 20 M88 E01072...
Page 76: ...Hardware Installation for the D4 ENDEAVOR V NTEC 1 Detector User Manual 5 10 M88 E01072...
Page 96: ...Calibration V NTEC 1 Detector User Manual 7 8 M88 E01072...
Page 102: ...Alignment V NTEC 1 Detector User Manual 8 6 M88 E01072...
Page 122: ...Appendix A Configuration Tables V NTEC 1 Detector User Manual A 4 M88 E01072...
Page 128: ...Appendix B Spare Parts List V NTEC 1 Detector User Manual B 6 M88 E01072...
Page 137: ...M88 E01072 D 1 Appendix D EC Declaration of Conformity See the figure on the next page...
Page 138: ...Appendix D EC Declaration of Conformity V NTEC 1 Detector User Manual D 2 M88 E01072...