Calibration
AD1200 Reference
iv)
Adjust the OFFSET trimpot so that the input code from the
A/D converter flickers evenly between 00hex and 01hex.
This is the OFFSET adjustment.
For BIPOLAR ranges adjust trim pot 2 for 00hex to 01hex
transition. eg for range -5 to +5 bipolar the 000+1/2LSB
voltage is -4.9988 Volt.
For UNIPOLAR ranges adjust trim pot 1 for 00hex to
01hex transition. eg for range 0 to +5 unipolar the
000+1/2LSB voltage is 610 uVolt.
v)
Using Figures 4-7 & 4-8 determine the precision voltage
source output for the 1 1/2 LSB code below FSR, the Full
Scale Reading, the FSR-1.5LSB value.
vi)
Adjust the GAIN trimpot so that the input code from the
A/D converter flickers evenly between FEhex and FFhex.
This is the GAIN adjustment.
Adjust trim pot 3 for for FEhex to FFhex transition.
eg for range -5 to +5 bipolar the FSR-1.5LSB value is
+4.9963 Volts.
eg for range 0 to +5 unipolar the FSR-1.5LSB value is
+4.9982 Volts.
Figure 4-7. Unipolar ADC Calibration Values.
________________________________________
ÚÄÄÄÄÄÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³
³
ANALOGUE INPUT RANGE
³
³
DIGITAL
ÃÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ´
³
OUTPUT
³
UNIPOLAR
³
³
CODE
ÃÄÄÄÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÄÄÄÂÄÄÄÄÄÄÄÄÄÄ´
³
³
0 to 1.25
³
0 to+2.5V
³
0 to +5V
³
0 to +10V
³
ÃÄÄÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄ´
³
FSR
³
+1.2500V
³
+2.5000V
³
+5.0000V
³
+10.0000V
³
³
FSR-1LSB
³
+1.2497V
³
+2.4994V
³
+4.9988V
³
+9.9976V
³
³
FSR-1.5LSB
³
+1.2495V
³
+2.4991V
³
+4.9982V
³
+9.9963V
³
ÃÄÄÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄÅÄÄÄÄÄÄÄÄÄÄ´
³
000 hex
³
0.0000V
³
0.0000V
³
0.0000V
³
0.0000V
³
³
1 LSB
³
0.305mV
³
0.61mV
³
1.22mV
³
2.44mV
³
³
1/2 LSB
³
0.153mV
³
0.305mV
³
0.61mV
³
1.22mV
³
³
000+1/2LSB
³
153uV
³
305uV
³
610uV
³
1220uV
³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÄÄÄÁÄÄÄÄÄÄÄÄÄÄÙ
Chapter 4
Page 79