Calibration
AD1200 Reference
Outline.
_______
The calibration is simple. It consists of only two steps.
i)
Set the D/A converter to its LOWEST output, then using
an accurate voltmeter check the output voltage, adjust the
OFFSET trim potentiometer to get the desired value.
ii)
Set the D/A converter to its HIGHEST output, then using
an accurate voltmeter check the output voltage, adjust the
GAIN trim potentiometer to get the desired value.
Provided the calibration is performed in the correct order,
OFFSET adjust then GAIN adjust, there should be no interaction
between the two adjustments.
Equipment Required.
___________________
A digital voltmeter capable of a stable and clear resolution
of 1/10th of LSB at both ends of the range of the converter.
Thus the voltmeter should have a range of -10 to +10 volts and
a resolution of 1.22mV/10 = 122uV or better.