AD1200 Reference
Calibration
FE to FF hex.
Provided the calibration is performed in the correct order,
OFFSET adjust then GAIN adjust, there should be no interaction
between the two adjustments.
Equipment Required.
___________________
A precision voltage source capable of a stable output and
resolution of 1/10th of LSB at both ends of the range of the
converter . Thus the voltage source should have a range of -5 to
+10 volts and a resolution of 1.22mV/10 = 122uV or better.
Figure 4-6. ADC Calibration Wiring Diagram.
________________________________________
ÚÄÄÄÄÄÄÄÄÄÄÄÄ¿
³
PRECISION
³
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³
VOLT SOURCE
³
³ÚÄÄ
³
ÚÄÄÄÄÄÄÄÄ¿
³
³³
³
ÀÄÄÄÄÄÄÄÄÙ
³
ÚÄij³Ä¿
³
+ve -ve
³
³
³³
Pin
ÀÄÄÂÄÄÄÄÄÄÂÄÄÙ
CH0 INPUT
ÄÄÄÄÄÄÄÄÄÄÄÄÄı±±ÄÄÄÄÄÄÄÄÄÄÄÄÄÙ
³
³
³³
Pin
³
CH8 ( =CH0 RET) INPUT
ÄÄÄÄÄÄÄÄÄÄÄÄÄı±±ÄÂÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÙ
³
³³
³
CH1-CH7, CH9-CH15
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³
³
³³
³
³
AGND
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÁÄı±±Ä´
³
³³
³
AMP LO
ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄı±±ÄÙ
ÀÄij³ÄÙ
ÄÄÄÄ¿
ÚÄÄÙ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÙ
³
ADC Calibration Procedure.
_________________________
i)
Set up the ADC input jumpers for the input range required,
make a note of the input range.
ii)
Using leads as short as possible connect up the digital
voltmeter as in Figure 4-5
iii)
Using Figures 4-7 & 4-8 determine the precision voltage
source output for the 1/2 LSB code above that producing
the 000 hex output code, the 000+1/2LSB value.
Page 78
Chapter 4