Digital Power Meter Display, 5000-NG
55
Measurement procedure
If additional information about any settings is desired prior to beginning a
measurement, see Additional Information.
1. Select Sensor Mode on the Sensor Operation Menu .
Average Power
Time Domain
Statistical Power
Average Power Measurement
2. Tap the Display Controls Menu.
3. Tap Configuration.
4. Enter Bandwidth value, if desired (see Bandwidth (Video Filter)).
5. Tap the Display Controls Menu.
6. Tap Readings.
Time Domain Measurement
2. Tap the Display Controls Menu.
3. Tap Configuration.
4. Select Bandwidth value, if desired (see Bandwidth (Video Filter)).
5. Enter Sample Interval.
6. Select Trigger source and associated values, if desired.
7. Tap the Display Controls Menu.
8. Tap Readings.
9. Tap the Sensor Operation Menu.
10. Tap Device Actions.
11. Tap Start to initiate measurement.
Statistical Power Measurement
2. Tap the Display Controls Menu
3. Tap Configuration.
4. Select Bandwidth value, if desired (see Bandwidth (Video Filter)).
5. Select normal Run Mode (single sweep) or Clear/Restart (continuous
sweeps).
6. Select Confidence Percentage, Sample Count, OR Sample Time. Change only
one of these parameters. See Confidence Percentage.
7. Enter the dynamic Range value. This sets the dB scale (x-axis) for the graph.