Veterinary monitor user’s manual
Chapter 5-27
Warning: Disposable pressure transducer should not be reused. And it must be
used before expired data. Do read the expired data on the IBP accessory package
bag.
Warning: When the monitor is used with HF surgical equipment, the
transducer and the cables must be avoided conductive connection to the HF
equipment to protect against burns to the patient.
The specified transducer is designed to protect against the effects of a discharge of a
cardiac defibrillator. When the patient is in the defibrillation, the waveform of IBP maybe
distorted temporarily. After the defibrillation, the monitoring will go on normally, the
operation mode and the user configuration are not affected.
Warning: The operator should avoid contact with the conductive parts of the
appurtenance when being connected or applied.
2) Plug the cable of IBP into the IBP socket on the right panel of the monitor. Connect the
extension tube of the transducer and blood vessel with the artery needles and secure them,
then make sure three-way valve 1 and three-way valve 2 are in a state of ON. At this
moment, BP waveforms should appear on the screen of the monitor.
3) Set up parameters and modes relevant to IBP.
5.6.3 Setup of IBP Parameters
IBP Label
Select the names of IBP labels. Options are
IBP1, IBP2, ART, CVP, PA,
RAP, ICP
and
LAP.
Unit
Select the units of IBP, and options are
mmHg, kPa
and
cmH
2
O
.