Veterinary monitor user’s manual
Chapter 6-9
upper alarm limit
IBP SYS1 too low
SYS measuring value of channel 1 is below
lower alarm limit
User-Selectable
IBP DIA1 too high
DIA measuring value of channel 1 is above
upper alarm limit
User-Selectable
IBP DIA1 too low
DIA measuring value of channel 1 is below
lower alarm limit
User-Selectable
IBP MAP1 too high
MAP measuring value of channel 1 is above
upper alarm limit
User-Selectable
IBP MAP1 too low
MAP measuring value of channel 1 is below
lower alarm limit
User-Selectable
IBP SYS2 too high
SYS measuring value of channel 2 is above
upper alarm limit
User-Selectable
IBP SYS2 too low
SYS measuring value of channel 2 is below
lower alarm limit
User-Selectable
IBP DIA2 too high
DIA measuring value of channel 2 is above
upper alarm limit
User-Selectable
IBP DIA2 too low
DIA measuring value of channel 2 is below
lower alarm limit
User-Selectable
IBP MAP2 too high
MAP measuring value of channel 2 is above
upper alarm limit
User-Selectable
IBP MAP2 too low
MAP measuring value of channel 2 is below
lower alarm limit
User-Selectable
Technical Alarm
Message
Cause
Alarm Level
IBP1 sensor off
IBP cable of channel 1 falls off from monitor
Low
IBP2 sensor off
IBP cable of channel 2 falls off from monitor
Low
IBP communication error
IBP communication error
Low
IBP1 alarm error
Alarm failure
Low
IBP2 alarm error
Alarm failure
Low
Prompt:
Message
Cause
Alarm Level
IBP1 Checking
IBP1 zero calibration is in progress.
No alarm
IBP1 Errlose
IBP1 zero calibration failed for IBP1 cable
falls off.
IBP1 Errtimeout
IBP1 zero calibration failed for time is out.
IBP1 Check OK
IBP1 zero calibration success.
IBP2 Checking
IBP2 zero calibration is in progress.
IBP2 Errlose
IBP2 zero calibration failed for IBP2 cable
falls off.
IBP2 Errtimeout
IBP2 zero calibration failed for time is out.
IBP2 Check OK
IBP2 zero calibration success.