Veterinary monitor user’s manual
Chapter 6-7
Technical Alarm
:
Message
Cause
Alarm Level
ECG RA LA LL V- LEAD
OFF
ECG electrode fall off the skin or ECG
cables fall off the monitor
Low
ECG electrode polarized
ECG electrode polarized
Low
ECG communication error
ECG measurement failure or communication
failure
Low
HR alarm error
Alarm failure
Low
6.8.2
RESP Alarm
Physiological Alarm
:
Message
Cause
Alarm Level
RR too high
RR measuring value is above the upper alarm
limit
User-Selectable
RR too low
RR measuring value is below the lower
alarm limit
User-Selectable
RESP Apnea
No signal for breath in specific interval
User-Selectable
Technical Alarm
:
Message
Cause
Alarm Level
RR alarm error
Alarm failure
Low
6.8.3
SpO
2
Alarm
Physiological Alarm
:
Message
Cause
Alarm Level
SpO
2
too high
SpO
2
measuring value is above the upper
alarm limit
Medium ,High
User-Selectable
SpO
2
too low
SpO
2
measuring value is below the lower
alarm limit
Medium ,High
User-Selectable
PR too high
PR measuring value is above the upper alarm
limit
User-Selectable
PR too low
PR measuring value is below the lower alarm
limit
User-Selectable
Technical Alarm
:
Message
Cause
Alarm Level
SpO
2
sensor off
SpO
2
sensor may be disconnected from the
patient or the monitor
Low
SpO
2
communication error
SpO
2
measurement failure or communication
error
Low
SpO
2
alarm error
Alarm failure
Low
PR alarm error
Alarm failure
Low