Veterinary monitor user’s manual
Chapter 2-2
Defibrillator-proof type CF equipment (refer to IEC 60601-2-27)
The unit displaying this symbol contains a F-Type isolated (floating)
applied part providing a high degree of protection against shock, and is
defibrillator-proof.
The type CF applied parts provide a higher degree of protection against
electric shock than that provided by type BF applied parts.
Attention! Please refer to the documents accompanying this monitor
(this manual)!
Defibrillator-proof type BF equipment (IEC 60601-1:1995)
The unit displaying this symbol contains a F-Type isolated (floating)
applied part providing a high degree of protection against shock, and is
defibrillator-proof.
4. When a defibrillator is applied on a patient, the monitor may have transient disorders in
the display of waveforms. If the electrodes are used and placed properly, the display of the
monitor will be restored within 10 seconds. During defibrillation, please note to remove
the electrode of chest lead and move the electrode of limb lead to the side of the limb. The
electrode of the defibrillator should not come into direct contact with the monitoring
electrodes. Please ensure the monitor is reliably grounded and the electrodes used
repeatedly should be kept clean.
Warning: When conducting defibrillation, do not come into contact with the
patient, the bed and the monitor. Otherwise serious injury or death could be resulted
in.
5. To guarantee the safe operation of the monitor, Monitor is provided with various
replaceable parts, accessories and consuming materials (such as sensors and their cables,
electrode pads). Please use the products provided or designated by the manufacturer.
6. Monitor only guarantees its safety and accuracy under the condition that it is connected
to the devices provided or designated by manufacturer. If the monitor is connected to other
undesignated electrical equipment or devices, safety hazards may occur for causes such as
the cumulating of the leakage current.
7. To guarantee the normal and safe operation of the monitor, a preventive check and
maintenance should be conducted for the monitor and its parts every 6-12 months