Veterinary monitor user’s manual
Chapter 6-13
Sensor Warm Up
Module is warming up.
No Alarm
6.8.10 AG alarm and promotion
Physiological alarm:
Message
Cause
Alarm Level
EtAA too high
EtAA is above upper alarm limit
User-Selectable
EtAA too low
EtAA is below lower alarm limit
FiAA too high
FiAA is above upper alarm limit
User-Selectable
FiAA too low
FiAA is below lower alarm limit
EtN
2
O too high
EtN
2
O is above upper alarm limit
User-Selectable
EtN
2
O too low
EtN
2
O is below lower alarm limit
FiN
2
O too high
FiN
2
O is above upper alarm limit
User-Selectable
FiN
2
O too low
FiN
2
O is below lower alarm limit
EtO
2
too high
EtO
2
is above upper alarm limit
User-Selectable
EtO
2
too low
EtO
2
is below lower alarm limit
FiO
2
too high
FiO
2
is above upper alarm limit
User-Selectable
FiO
2
too low
FiO
2
is below lower alarm limit
Technical Alarm
:
Message
Cause
Alarm Level
GAS communication error
GAS module failure or communication error
Medium
Check Airway Adapter
Airway adaptor of GAS module
disconnected with sensor
Medium
Replace O
2
sensor
Oxygen sensor disconnected with module
Medium
O
2
sensor low
Weak oxygen sensor signal
Medium
GAS sensor error
GAS sensor error
Low
GAS CONC. Out of Range
Measurement of GAS module over range
Medium
Room Air Calibration
Required
Measurement of oxygen density is not
correct.
High