Veterinary monitor user’s manual
Chapter 6-4
Alarm recording configuration
Alarm indication of physiological parameters
Audio: when alarm occurs, the system generates alarm sound to raise the user’s
attention (audio alarm can be disabled).
Visual: The parameter flashes on the display area of the screen and alarm LED lights.
Warning
:
The lower limit and the upper limit of parameter must be set based on
clinical practices and general clinical experiences.
Note
:
When parameter alarm level is off, alarm will be disabled, even if the
measurement results exceed the limits. Alarm indicating lamp in the front of the
monitor will alarm at the highest level, if different levels alarms coexist.
6.4 Alarm Cause
Alarm of the monitor includes:
1. Physiological Alarm
2. Technical Alarm
3. General Prompt
Physiological Alarm
When the measuring value has exceeded the set parameter limit and its <ALM LEV>
is not <OFF>, the monitor alarms. The monitor wouldn’t alarm with absence of either
of the two conditions.