Veterinary monitor user’s manual
Chapter 1-17
Alarm indication
Sound and light indication
EtCO2 (Microstream
,
LoFlo)
Measure method
Infrared spectrum
Measure mode
Microstream
Warm up time
Capnogram displayed in less than 20 s, At an
ambient temperature of 25
℃
, full specifications
within 2 minutes.
CO
2
Measurement Range
0 ~ 19.7%(0 ~ 150 mmHg)
CO
2
Resolution
1mmHg
CO
2
Stability
Short-Term Drift: Drift over four hours
≤
0.8mmHg.
Long-Term Drift: Accuracy specification will be
maintained over a 120 hours period.
unit
%, mmHg, kPa
CO
2
Accuracy
(at 760 mmHg, ambient
temperature of 25°C)
0 ~ 40 mmHg, ±2 mmHg
41 ~ 70 mmHg, ±5% of reading
71 ~100 mmHg, ±8% of reading
101 ~ 150 mmHg, ±10% of reading
Above 80 breath per minute ± 12% of reading
Gas temperature at 25
℃
.
CO
2
response time
<
3s (includes transport time and rise time)
Respiration Rate Range
2~150 rpm
Respiration Rate Accuracy
±1 rpm
Sample Flow Rate
50 ml/min ±10 ml/min
Alarm indication
Sound and light indication
EtCO2 (Mainstream
,
CAPNOSTAT5)
Measure method
Infrared spectrum
Measure mode
Mainstream
Warm up time
Capnogram displayed in less than 15 s, At an
ambient temperature of 25
℃
, full specifications
within 2 minutes.
CO
2
Measurement Range
0 ~ 19.7%(0 ~ 150 mmHg)
CO
2
Resolution
1mmHg
CO
2
Accuracy
0 ~ 40 mmHg, ±2 mmHg
41 ~ 70 mmHg, ±5% of reading
71 ~100 mmHg, ±8% of reading
101 ~ 150 mmHg, ±10% of reading