Veterinary monitor user’s manual
Chapter 2-4
TO CLEAR PATIENT DATA
When monitoring a new patient, you must clear all previous patient data from the
system. To accomplish this, shut down the device, then turn on it. Selecting
〈
New
patient
〉
in
〈
main setup
〉
menu can also clear the previous patient data .
DISPOSAL OF PACKAGE
Dispose of the packaging materials, please observe the applicable waste control
regulations and keeping it out of children’s reach.
EXPLOSION HAZARD
Do not use this equipment in the presence of flammable anesthetics, vapors or liquids.
LEAKAGE CURRENT TEST
When interfacing with other equipment, a test for leakage current must be performed
by qualified biomedical engineering personnel before using with patients.
BATTERY POWER
The device is equipped with a battery pack. The battery discharges even when the
device is not in use. Store the device with a fully charged battery and take out the battery,
so that the service life of the battery will not be shortened.
DISPOSAL OF ACCESSORIES AND DEVICE
Disposable devices are intended for single use only. They should not be reused as
performance could degrade or contamination could occur.
The service life of this monitor is five years. At the end of its service life, the product
described in this manual, as well as its accessories, must be disposed of in compliance
with the guidelines regulating the disposal of such products. If you have questions
concerning disposal of products, please contact manufacturer or its representatives.
EMC
Magnetic and electrical fields are capable of interfering with the proper performance
of the device. For this reason, make sure that all external devices operated in the vicinity
of the monitor comply with the relevant EMC requirements. X-ray equipment or MRI
devices are a possible source of interference as they may emit higher levels of
electromagnetic radiation. Also, keep cellular phones or other telecommunication
equipment away from the monitor.
INSTRUCTION FOR USE
For continuous safe use of this equipment, it is necessary that listed instructions were
followed. However, instructions listed in this manual in no way can supersede established
medical practices concerning patient care.
LOSS OF DATA
Should the monitor at any time temporarily lose patient data, close patient observation
or alternative monitoring devices should be used until monitor function is restored.
If the monitor does not automatically resume operation within 60 seconds, restart the
monitor using the power on/off switch. Once monitoring is restored, you should verify
correct monitoring state and alarm function.