Veterinary monitor user’s manual
Chapter 5-6
ECG alarm
Click and open the dialog of HR alarm
Adjust alarm
Select this
option
to enter the configuration of alarm limits and configure
the limits by turning the trim knob to select the high limits and low limits, and exit by
selecting
〈
EXIT
〉
. The upper part is the high limit and the lower part is the low limit.
The configuration range of high limit is
0
~
350 bpm
continuously adjustable, not lower
than low limit and the configuration range of low limit is
0
~
350 bpm
continuously
adjustable, not higher than the high limit.
HR alarm
Select <
ON
> to enable HR over limit alarm; select <
OFF
> to disable HR
over limit alarm.
ECG Setup
Lead Type
Select the lead type of ECG input, and options are
5 leads, 3 leads, Auto
.
Scan speed
Select the scanning speed of ECG waveforms and options are
12.5mm/s,