Veterinary monitor user’s manual
Chapter 5-17
Use a piece of clean cloth to wipe the surface of the cable with 70% isopropyl alcohol,
a 10% bleach solution or 2% Cidex®, clean with clear water and wipe it dry.
Warning: Disposable TEMP probes must not be re-sterilized or reused.
Note: For protecting environment, the disposable TEMP probe must be recycled
or disposed of properly.
Disposal Note: Should the TEMP probe become damaged beyond repair, or for
some reason its useful life is considered to be at an end, please observe all local, state,
and federal regulations that relate to the disposal of this products or similar
products.
Warning: The calibration of temperature measurement is necessary for every
two years (or as frequently as dictated by your Hospital Procedures Policy). When
you need calibrate the temperature, contact the manufacture please.
Note: The self-test of the temperature measurement is performed automatically
once every 10 minutes during the monitoring. The test procedure lasts about one
second and does not affect the normal measurement of the temperature monitoring.
Note: If Temperature to be measured beyond probe’s measuring range, over
measuring range alarm will display on the screen. Check out if probe is on the
corresponding patient body site, or change it to other site on the patient.
Note: If “TEMP self-check error” display on the screen, it is possibly that
something is wrong with the temperature capture circuit, the operator should stop
using the monitor and contact with the company.
5.5 Measurement of NIBP
5.5.1 Brief Introduction to Measurement of NIBP
The monitor automatically conducts measurement of NIBP with the method of
shockwave. The method of shockwave indirectly estimates the systolic and diastolic
pressures within the blood vessels by measuring the change of the pressure within blood
pressure cuff along with the volume of the arteries and calculates the average pressure.