Monitor the CO2 waveform (Capnogram) for elevated baseline. Elevated baseline
can be caused by sensor or patient problems.
Periodically check the CAPNOSTAT5 sensor and tubing for excessive moisture
or secretion buildup.
Do not operate the CAPNOSTAT5 sensor when it is wet or has exterior
condensation.
Caution:
Use only accessories provided by manufacturer.
Do not sterilize or immerse the CAPNOSTAT5 sensor in liquids.
Do not clean the CAPNOSTAT5 sensor and accessories except as directed in this
manual.
It is recommended that the CO
2
sensor be removed from the circuit whenever an
aerosolized medication is delivered. This is due to the increased viscosity of the
medications which may contaminate the sensor windows, causing the sensor to
fail prematurely.
Do not apply excessive tension to the CAPNOSTAT5 sensor cable.
Note:
This product and its accessories are latex free.
After the life cycle of the CAPNOSTAT5 sensor and its accessories have been met,
disposal should be accomplished following national and local requirements.
Nitrous oxide, elevated levels of oxygen and helium can influence the CO2
measurement. Please setup gas compensation according to actual state.
Barometric pressure compensation is required to meet the stated accuracy of the
CAPNOSTAT5 sensor.
Do not place the combined CO2 sensor between the ET tube and the elbow, as
this may allow patient secretions to block the adapter windows.
Position the combined CO2 sensor with its windows in a vertical and not a
horizontal position: this helps keep patient secretions from pooling on the
windows.
5
.10.5 Maintenance and Cleaning
For CAPNOSTAT 5 Sensor and LoFlo Module
The outside of the module or sensor may be cleaned and disinfected by wiping with
70% isopropyl alcohol, a 10% bleach solution, or mild soap. After cleaning, wipe with a
clean, water-dampened cloth to rinse. Dry before use.
For Reusable Airway Adapters
Reusable airway adapters may be cleaned by rinsing in a warm soapy solution, followed
by soaking in a liquid disinfectant such as 70% isopropyl alcohol, a 10% bleach solution,