Chapter 5 Setting System Parameters
-- 38 -- User Manual for Electrocardiograph
Name
Value
Default
Description
will be displayed.
Rhythm Lead
I, II, III, aVR, aVL, aVF, V
II
Select a lead as rhythm lead.
Rhythm Time
30 s
~
300 s
60 s
Select one option for rhythm recording time.
Pacemaker
Detection
Disable, Weak, Normal,
Enhance
Disable
Select one option for the pacemaker.
Arrhythmia Mode
Disable, Enable
Disable
Set the arrhythmia mode.
Wave Sample Time
10 s
~
24 s
10s
Select one option for waveform sampling
time.
Pre-acquisition
Time
0s~10s
8s
Set the pre-acquisition time.
Printout Duration
3~12s
3s
Set the printout duration. Note that this
option is only available for iE 101.
Pre-acquisition
Disable, Enable
Disable
Set the pre-acquisition mode.
If enabled, it can print out previous
waveforms.
Auto-save
Disable, Enable
Enable
Set whether to automatically save the report.
Data Format
ECG, JPEG, PDF
ECG
The data format for storage.
Auto Upload
Disable, Enable
Disable
Set to upload the waveforms and reports
automatically or not after printing.
QTC Formula
Bazett, Fridercia,
Framingham, Hodges
Hodges
Select one option for QTC formula.
Examination Type
Normal, Physical
Examination
Normal
Select one option for examination type.
EXPLANATION
For physical examination of a large population, it is recommended to select
[Examination Type]
>
[Physical
Examination]
.
Summary of Contents for iE 101
Page 2: ......
Page 4: ......
Page 10: ...Contents IV User Manual for Electrocardiograph Blank Page ...
Page 18: ...Foreword 8 User Manual for Electrocardiograph Blank Page ...
Page 26: ...Chapter 1 Introduction 16 User Manual for Electrocardiograph Blank Page ...
Page 54: ...Chapter 5 Setting System Parameters 44 User Manual for Electrocardiograph Blank Page ...
Page 74: ...Appendix B Technical Specifications 64 User Manual for Electrocardiograph Blank Page ...
Page 81: ......