Appendix B Technical Specification
User Manual for Electrocardiograph -- 61 --
Appendix B Technical Specification
B.1 Specifications
B.1.1 Main Unit
Lead
7-lead
Acquisition Mode
Simultaneous 7-lead Acquisition
Record Format
1
×
7, 1
×
7+1R (iE 101)
3
×
2+1, 3
×
2+1+1R, 3+4 (iE 300)
Record Mode
Auto, Manual, Upload
Lead Format
7
×
1, 3+4
Rhythm Time
30~300s waveforms acquisition for rhythm analysis
Measurement
Parameters
Ventricular Rate, PR Interval, QRS Time Limit, QT/QTC Interval, P/QRS/T Axis
Filters
AC Filter
Baseline Wander Filter
Low-pass Filter
Input CIR Current
≤0.1 μA
Input Impedance
≥30 M
(10Hz)
Time Constant
≥3.2 s
Frequency Response
0.01 Hz
~
250 Hz
Noise Level
≤12.5 μV
p-p
Sensitivity Threshold
≤20 μV
p-p
Sensitivity
Auto, 0.625 mm/mV, 1.25 mm/mV, 2.5 mm/mV, 5 mm/mV, 10/5 mm/mV, 10
mm/mV, 20/10mm/mV, 20 mm/mV, and 40 mm/mV
Standard Sensitivity
10 mm/mV ± 2%
Calibration Voltage
1 mV±3 %
Accuracy
of
input
signal reproduction
Using the method described in 4.2.7.1 of AAMI EC11 to test the overall system error,
which is within ±5%;
Summary of Contents for iE 101
Page 2: ......
Page 4: ......
Page 10: ...Contents IV User Manual for Electrocardiograph Blank Page ...
Page 18: ...Foreword 8 User Manual for Electrocardiograph Blank Page ...
Page 26: ...Chapter 1 Introduction 16 User Manual for Electrocardiograph Blank Page ...
Page 54: ...Chapter 5 Setting System Parameters 44 User Manual for Electrocardiograph Blank Page ...
Page 74: ...Appendix B Technical Specifications 64 User Manual for Electrocardiograph Blank Page ...
Page 81: ......