Chapter 4 ECG Recording
-- 32 -- User Manual for Electrocardiograph
4.3.1 Auto Mode
In Auto Mode, the ECG machine can automatically print the waveforms and reports.
[Report Format]
The ECG machine can analyze the resting ECG and output the measurement data, median beat, etc.
Simple Report
includes animal information, simple measurement data;
Median Beat Report
includes animal information, simple measurement data, Median Beat waveform and
Rhythm waveforms.
In Manual Mode, the analysis reports include analysis report (1), and analysis report (2).
After printing waveforms and report, press
to print waveforms and report of the last animal.
Auto Print
In Automatic mode, select
[Menu] > [Print Setting] > [Auto Print] > [Enable]
, the ECG machine wil
automatically print the waveforms and reports when the following 3 conditions are satisfied.
Lead off does not detected for 2s (none of any electrodes);
Five or more QRS complex detected;
Waveform is stable, no EMG interference or Baseline drift.
EXPLANATION
The ECG machine analyzes the latest 10 seconds waveforms.
Not print
Print the selected
report
Edit and then
Disable
Simple Report/ Median Beat
Report
Manual Report
Disable
Simple Report
Median Beat Report
Manual Report
Summary of Contents for iE 101
Page 2: ......
Page 4: ......
Page 10: ...Contents IV User Manual for Electrocardiograph Blank Page ...
Page 18: ...Foreword 8 User Manual for Electrocardiograph Blank Page ...
Page 26: ...Chapter 1 Introduction 16 User Manual for Electrocardiograph Blank Page ...
Page 54: ...Chapter 5 Setting System Parameters 44 User Manual for Electrocardiograph Blank Page ...
Page 74: ...Appendix B Technical Specifications 64 User Manual for Electrocardiograph Blank Page ...
Page 81: ......