Chapter 5 Setting System Parameters
User Manual for Electrocardiograph -- 37 --
Chapter 5 Setting System Parameters
In the Main interface, select
[Menu]
to enter the configuration menu.
Figure 5- 1 Configuration Menu
Press
to return to the previous page; press
step by step to return to the Main interface.
Select
[Save]
before exiting to avoid loss of settings caused by sudden power loss.
5.1 ECG Setting
Enter
[ECG Setting]
to adjust parameters about Electrocardiograph. See following table:
Name
Value
Default
Description
Low-pass Filter
25 Hz, 35 Hz, 75 Hz, 100
Hz, 150 Hz, 250 Hz
100 Hz
Select one option for low-pass Filter,
Baseline Wander Filter and AC Filter.
Baseline
Wander
Filter
0.01 Hz, 0.02 Hz, 0.05 Hz,
0.35 Hz, 0.5 Hz, 0.8 Hz
0.5 Hz
AC Filter
OFF, 50 Hz, 60 Hz
50 Hz
V Lead
Disable, Enable
Enable
When it is enabled, 7 waveforms will be
displayed; when it is disabled, 6 waveforms