Chapter 7 Maintenance
-- 52 -- User Manual for Electrocardiograph
Figure 7- 1 Battery Replacement
EXPLANATION
Refer to
Foreword
for other warning information of battery.
7.6 Silicon Rubber Shaft for Printing
The silicon rubber shaft shall be kept clean, smooth and free from dirt. Otherwise, the ECG machine may print out
unsatisfied ECG trace. To clean the shaft, wipe the shaft with soft cotton moistened with medical alcohol and at
the same time rotate shaft until it is clear enough.
7.7 Thermal Print Head
Residue and dirt on the thermal print head could affect the clarity of recorded ECG waveform. To clean the
thermal print head, open the paper cover and clean the print head with soft cotton moistened with medial alcohol.
It is not permitted to clean the print head with a sharp object, which can cause permanent damage to the print head.
Thermal print head maintenance should be done at least once a month.