Chapter 2 Preparation
-- 22 -- User Manual for Electrocardiograph
If the examination involves a short period of time, if ECG gel is unavailable, please wipe the skin with
medical alcohol to keep the skin clean and moist, rapidly attach the electrodes.
It’s not allowed to use saline water as substitute when ECG gel is not available. The saline water will cause
corrosion on the electrodes.
Electrodes shall be properly stored. When electrodes have been used for a certain period, they may become
corroded and oxidized at the surface. Whenever this happens, the electrodes must be replaced.
Do not mix electrodes of different types and manufacturers. Do not use re-useable electrodes and disposable
ones together, or it will affect the recording.
Please use our company’s or authorized electrodes to make sure qualified ECG signals.
2.7.1 Electrodes Attachment
Limb Electrodes Placement
IEC
AHA
Description
Figure
R Red
RA White
Right Arm
L Yellow
LA Black
Left Arm
N Black
RL Green
Right Leg
F Green
LL Red
Left Leg