Chapter 2 Preparation
-- 20 -- User Manual for Electrocardiograph
2.5 Power On/Off
Press
to power on/off the ECG machine.
The ECG machine enters Standby mode if it is not in use for a setting period. Set the duration in
[System
Setting] > [Standby Time]
. To exit standby mode, press any key.
The ECG machine will shut down automatically if it is not in use for a setting period. Set the duration in
[System
Setting] > [Auto Power Off]
.
2.6 Connect to Network
CAUTION
In the data transmission, if the ECG machine warns “Network connection failed”, please reconnect to network or
reset the network configuration.
1.
As shown in the following figure, the cable network system is composed of the ECG machine, the
switchboard and the server.
2.
Go to
[Menu]
>
[System Setting] >
[Cable Network]
and set the
[IP address]
,
[Subnet Mask]
and
[Default Gateway]
of the ECG machine. If the IP address is in the same network segment of the server, the
subnet mask and gateway shall be as the set value of the server. If the IP address is not within the same
network segment of the server, subnet mask and gateway of the
[Cable Network]
shall be set according to
actual situations. And make sure the specified gateway does support the data transmission between the two
network segments.
3.
Select
[Server Setting]
and set the correct
[IP address]
and
[Port]
number of the server.
4.
When the machine is connected to network successfully, the icon
will display in the main interface.
Summary of Contents for iE 101
Page 2: ......
Page 4: ......
Page 10: ...Contents IV User Manual for Electrocardiograph Blank Page ...
Page 18: ...Foreword 8 User Manual for Electrocardiograph Blank Page ...
Page 26: ...Chapter 1 Introduction 16 User Manual for Electrocardiograph Blank Page ...
Page 54: ...Chapter 5 Setting System Parameters 44 User Manual for Electrocardiograph Blank Page ...
Page 74: ...Appendix B Technical Specifications 64 User Manual for Electrocardiograph Blank Page ...
Page 81: ......