Foreword
-- 2 -- User Manual for Electrocardiograph
defibrillation polarization recovery time (however, they are essential accessories of ECG), should not be used
immediately for reliable measurements and diagnostics after defibrillation. To ensure proper defibrillator
protection, use only the recommended disposable electrodes (Name: Skintact; Type, RT-34), lead wire and
electrode adapters by our company. To ensure the protection of defibrillator discharge, use the lead wire with
defibrillation-proof by our company
.
When the ECG machine is used together with a defibrillator or other electrical stimulators (like
high-frequency surgical devices), we recommend using disposable chest electrodes. Otherwise the animal
may get serious injury by using mental electrodes.
During defibrillation, the device can detect the discharge of defibrillator, and process automatically, and then
quickly recover the waveform.
Keep the ECG machine electrodes away from the electrodes of high-frequency electrosurgical units. Ensure
the resistance between the electrosurgical unit and animal body is as low as possible. If necessary, the
disposable electrodes can be used because of its larger contact area on the animal body, which can keep the
high-frequency current density in an acceptable range.
When the relevant packaging material including, depleted batteries and scrapping products are disposed,
please follow the local laws; the user should properly follow requirements of local laws, and recycling laws.
CAUTION
Avoid contacting with water or other liquids, and avoid using and storing in spaces with too large barometric
pressure, humidity and temperature beyond the prescribed standards, poor ventilation, or with excessive dust.
The ECG machine should be place on the flat horizontal table and avoid strong vibration and mechanical
shock while moving.
The frequency of AC power supply and system voltage should be complied with the requirements. More
importantly, the current capacity should be sufficient.
The instrument should not be surrounded by high-voltage cables, ultrasound equipment, electrotherapy
machines and other high-power equipment.
To more accurately record the ECG, the equipment shall be placed in a quiet and comfortable environment.
The circuit of application parts works based on floating ground and meets the safety standards in IEC60601-1
CF Type. It can be used in acquiring the body surface ECG signals, but cannot applied to the heart directly.
Turn off the ECG machine if an accident happens.
Summary of Contents for iE 101
Page 2: ......
Page 4: ......
Page 10: ...Contents IV User Manual for Electrocardiograph Blank Page ...
Page 18: ...Foreword 8 User Manual for Electrocardiograph Blank Page ...
Page 26: ...Chapter 1 Introduction 16 User Manual for Electrocardiograph Blank Page ...
Page 54: ...Chapter 5 Setting System Parameters 44 User Manual for Electrocardiograph Blank Page ...
Page 74: ...Appendix B Technical Specifications 64 User Manual for Electrocardiograph Blank Page ...
Page 81: ......