7
Installation Site Requirements
Calibration and performance verifications are conducted prior to shipping. Biocare Medical field service personnel will unpack or
supervise the unpacking of the ONCORE. Biocare field service personnel will verify the system performance. Biocare personnel will also
provide basic operating instructions.
Note: The ONCORE Automated Slide Staining Instrument must be handled using the bottom of the side chassis by at least two people
capable of lifting 75 pounds each. Do not lift the ONCORE Instrument from the front access cover.
The ONCORE Automated Slide Stainer must be placed on a solid, vibration-free laboratory bench with a flat area capable of safely
supporting the weight of the instrument. Additional consideration must be made for the desktop computer, monitor, printers, a minimum
clearance of 5.0 cm (2 in.) on all sides for ventilation and 40.64 cm (16 in.) above the top surface of the instrument to accommodate
the door in the “up” or “open” position.
The label and report printers for the ONCORE Automated Slide Stainer may be located anywhere that allows them to be connected to
the ONCORE Automated Slide Stainer’s dedicated PC. Surge protection is required. The backup uninterruptible power supply (UPS)
should be placed on a surface capable of safely supporting its weight, away from liquids and within 1 meter of the instrument power
cord and within 1 meter of a properly grounded power outlet. The use of a dedicated circuit is recommended for the instrument. Avoid
connecting the instrument on the same circuit as refrigeration equipment.
Environmental Conditions
Placement
Indoors
Operating temperature range
15-30°C (59-86°F). Recommended temperature for optimal staining 68-75°F
Maximum programmable heater temperature
110°C (230°F). See Protocol Editor Section for recommendations.
Maximum relative humidity
80%, non-condensing.
Ventilation
Do not block the vent openings on the instrument. Leave a minimum of 5 cm (2 in.) of space between the ONCORE Automated Slide
Stainer and the wall, cabinets or other obstructions.
Venting (exhaust)
The instrument does not need special exhaust ventilation unless required by safety regulations when using certain reagents.
Humidity
Avoid humidity extremes. Low humidity can cause electrostatic problems that may affect the computer. High humidity may create
condensation.
Temperature
The ONCORE Automated Slide Stainer must be placed in a location away from direct sunlight and extreme heat or cold.