71
FLYNET 100
5.5
Transmission
The indicator is equipped with external serial and Ethernet communication ports. It is therefore
possible to interface and communicate with various external devices such as displays, weight
repeaters, PCs, PLCs and so on.
The data that can be transmitted are linked to the indicator internal programme and therefore cannot
be changed.
The transmission configurations available are:
•
Mpp
•
Network
•
Serial
•
Field bus
5.5.1
Mpp
The Mpp or permanent weight memory is a legal storage system for weighing operations made with
the indicator. This system allows saving metrological data on the indicator and storing them for a
certain period of time strictly related to metrological laws. The minimum storage period for these
weighs is 180 days, after this time the indicator will allow the user to delete obsolete Mpps to have
more space on the disc. The amount of Mpp data that can be stored is directly proportional to the
size of the remaining memory on the internal disc. It is advisable to delete the obsolete data to avoid
unnecessary waste of memory, and to run into delays due to the processing of a huge amount of
data.
The Mpp is always present on the indicator, if the system is legalised, the weighing data are saved in
a table that can be consulted from the indicator. The data are stored as follows:
•
At each weight acquisition an identification code is generated
•
The identification code is linked to the acquisition time and date and the indicator number
•
The data are saved in a storage table
•
To access this table simply follow the following path: menu/info/MPP
•
To search for data, enter the date and click on the magnifying glass on the right side
•
If data are present, a code list is displayed
•
To view the data, simply click on the code displayed
•
A new page opens, which provides a summary of the stored weighing
The identification code consists of 7 digits, so the same code can only be reused after 10,000,000
weighing operations.
The MPP transmission allows sending a certified and legal data to the PC, as it is linked to the
indicator by an identification code. To activate the transmission of an Mpp string to a PC, the
accessory service must be activated. The activation of the Mpp service on the indicator allows using
the Mpp string, its remote controls and the display of Mpp data as they are acquired.