38
FLYNET 100
•
1:
the indicator has a single weighing range for the entire
range of the scale capacity.
•
2MD:
the indicator has two weighing ranges with MULTI-
DIVISION mode
1
.
•
2ME:
the indicator has two weighing ranges with MULTI-
EXTENSION mode
2
.
•
3MD:
the indicator has three weighing ranges with MULTI-
DIVISION mode.
•
3ME:
the indicator has three weighing ranges with MULTI-
EXTENSION mode.
1st range capacity
Value of the capacity corresponding to the first partial weighing
range.
2nd range capacity
Value of the capacity corresponding to the second partial
weighing range. (in case of 3 ME or 3 MD)
W1 extension return delay (no.
of conversions)
Number of samplings of delay on return to extension W1 for
multi-extension scales.
Manual range management
Possibility of selecting the ranges manually.
SAMPLING DATA
In calibration/sampling data it is possible to view and/or edit the sampling data. The first column
contains the real weight values and the second column contains the corresponding points.
(Empty Scale) Zero
Empty scale
(Maximum capacity) Full scale
Capacity
1st weight/loading intermediate
points
Calibration of 1st ascending load of the linearised calibration.
2nd weight/loading intermediate
points
Calibration of 2nd ascending load of the linearised calibration.
2nd weight/unloading
intermediate points
Calibration of 2nd descending load of the linearised calibration.
1st weight/unloading
intermediate points
Calibration of 1st descending unload of the linearised
calibration.
1
Multi-division = means that to use again the division of W1 it is mandatory to pass through 0
2
Multi-extension = means that to use again the division of W1 it is not mandatory to pass through 0