STABILITY
Number of
Stability
Divisions
0
Allows setting the number of divisions by which the weight may
vary while the weight stable signal is maintained. On indicators
subject to metrical verification, this parameter must be set to 0
(zero) or 1.
1
2
4
Stability Signal
Very fast (5
e)
Speed at which the stable weight signal is displayed. Indicates
how many equal consecutive readings are necessary to obtain
the stable weight signal. These consecutive readings may vary
according to the number of divisions indicated in the previous
field. This parameter is also related to the setting of the digital
filter parameter.
Fast
(10 e)
Slow
(20 e)
Very slow
(40 e)
Digital Filter
Allows adding or not a digital filter. It actually enables the addition or not
of a software processing for filtering the output values from the
analogue/digital converter. The higher the number, the greater the
filtering.
No
4
8
16
32
64
128